DISTILLATION UNIT – TECHNICAL SPECIFICATIONS

1. General features

  • Stainless steel body.
  • Programmable microprocessor control system.
  • Manual or automatic operation according to user preference.
  • All operations displayed on an alphanumeric LCD screen.

2. Functional features

  • Distillation and water addition are controlled automatically.
  • Continuous monitoring with a water level sensor; automatic warning when the level drops.
  • Distillation time adjustable from 0 to 49 minutes.
  • Automatic drain system empties waste liquids at the end of distillation.
  • Steam generator power of at least 1700 W.

3. Safety and warning systems

  • The user is informed whether there is water in the steam generator.
  • Error and alarm system covers: low water flow to the steam generator; safety breach in the absorption or digestion system; water level detection error.
  • The instrument shuts down automatically in the event of any fault or safety risk.

4. Chemical dosing control

  • The amount of alkali is adjustable and can be added by the user when needed.

DIGESTION UNIT – TECHNICAL SPECIFICATIONS

1. General features

  • Stainless steel body.
  • Processes 8, 12 or 24 samples at the same time.
  • Digestion temperature adjustable up to 450 °C.
  • One-piece fume exhaust system.

2. Sample equipment

  • Works with 300 ml digestion tubes, 50 mm in diameter.

TECHNICAL PARAMETERS

ParameterValue / Description
Sample typesGrain, feed, dairy products, beverages, soil, pharmaceuticals, sediments, chemicals, etc.
Operating modeAutomatic and manual
Water supplyTap water or distilled water
Sample capacitySolid: 0.20–2.00 g / Semi-solid: 2.00–5.00 g / Liquid: 10.00–25.00 ml
Measuring range0.1–200 mg N
Recovery rate≥ 99 %
Distillation time5–15 min/sample (depending on sample quantity)
Cooling water consumptionapprox. 3 L/min
RepeatabilityRelative standard deviation ≤ 1.5 %